Features:
• Multiple scales
• Temperature display
• Optical density display
• Temperature compensation
• LED/Interference filter
• Parallel printer port
• RS232 x 2
(refractometer/PC/barcode)
• Barcode reader
• High resolution LCD display
• Conforms to USP/EP/BP/JP
Software Features:
• GLP software (date/time/batch)
• 21CFR Part 11 audit trail
• Calibration audit trail
• Purity methods (Saccharimeter)
• Specific rotation methods
• Statistics
• Over 700 stored readings
Optional Extras:
• Funnel & flow tubes
• Low volume tubes
• Water-jacketed tubes
• Waterbaths
• Slotted lids
• Quartz control plates
ADP440 Polarimeter
The ADP440 is single wavelength, high accuracy polarimeter suitable for
use in
many applications and is especially suited for use in pharmaceutical laboratories
where compliance with Pharmacopoeia¹ is required. New optics featuring
a ‘no maintenance’ yellow LED and interference
filter together with a photodiode detector allows reading of samples of
up to 3.0 OD. The instrument has both angular and sugar scales (ISS) and
can be programmed with user scales as well as industry standard methods
for displaying concentration and specific rotation directly.
ADP440 - Operation
As the instrument is continually reading, all that is required is for the
operator to place the sample tube in the measurement chamber and read the
result. Four push buttons on the instrument keypad allow for MENU
setup, ZERO calibration, METHOD selection (or display MODE if
no methods are being used) and PRINT. A security system using passwords
protects against deliberate or accidental misuse, which can be expanded
to allow conformance with the technical requirements of FDA regulation
21 CFR Part 11.
The ADP440 continually displays the measured reading, providing the operator assurance of measurement and sample stability. Data handling is in accordance with GLP providing a record of measurement date, time, batch and operator numbers.
On-screen menus in English, French, Spanish or German are selectable.
Where sample temperature is critical, optional polarimeter tubes fitted with a water jacket may be used in place of standard glass tubes. The instrument temperature sensor may be placed in the centre arm of the tube and if required, sugar or a user defined temperature compensation may be adopted.
ADP440 – Sugar Purity Method
When used in the sugar industry, the ADP440 may also form the central hub
of a purity system, providing the measurement of a Saccharimeter. In the
‘purity mode’ the instrument will calculate and display ‘sugar
purity’ with Brix
values being input automatically from an RFM² refractometer or via
the simple
keypad. For rapid measurements, funnel and flow through cells are available
and require the use of specially suited slotted lids.
ADP440 - Calibration
The instrument can be checked and calibrated using a quartz control plate
by
accessing the span facility in the setup menu. If a sample is too dark,
the
instrument will display an error message.
1. Minimum requirement of EP/BP/JP/USP. 21CFR Part II
2. RFM700/340/800 only
|
|
|